What are the different results displayed on the HQ series meters?

Document ID

Document ID TE12914

Published Date

Published Date 12/20/2021
Print
Question
What are the different results displayed on the HQ series meters?
Summary
Explanation for the different readings.
Answer
With the HQ Series meters the temperature and raw measured values are displayed on the right side of the screen, and the calculated result is displayed in the center of the screen.

In the case of Conductivity, the raw non-temperature corrected measured result is displayed along with the measured temperature on the right. The temperature corrected result is displayed as the primary result in the center of the screen. For TDS, Salinity, or Resistivity, the raw non-temperature corrected measured Conductivity is displayed along with the measured temperature on the right and the TDS, Salinity, or Resistivity with temperature correction is displayed as the primary result in the center of the screen.

In the case of pH, the raw mV measurement is displayed along with the measured temperature on the right. The pH result calculated by calibration is displayed as the primary result in the center of the screen.

In the case of ORP, the raw mV measurement is displayed along with the measured temperature on the right. The mV result corrected by calibration is displayed as the primary result in the center of the screen.

In the case of LDO, the measured temperature, Pressure, and percent saturation is displayed on the right along with salinity when salinity correction is use. The O2 result by calibration is displayed as the primary result in the center of the screen (with salinity correction when applicable).

In the case of ISEs, the raw mV measurement is displayed along with the measured temperature on the right. The ppm result calculated by calibration is displayed as the primary result in the center of the screen.