PIB 37-652 Incorrect Labels on PN 1438901 (1.6N H2SO4) Digital Titrator Cartridges

Document ID

Document ID PIB529

Published Date

Published Date 05/09/2022
Title
PIB 37-652 Incorrect Labels on PN 1438901 (1.6N H2SO4) Digital Titrator Cartridges
Introduction
Product Update for distribution to Global subscribers.
Applies to the Hach Lab brands.

A label printing error has been identified on Sulfuric Acid 1.6N Digital Titrator Cartridges item Number 1438901

Key Message Points
  • Labels on some lots of item number 1438901 (Sulfuric acid,1.600N, cartridge) contain incorrect information.
  • The incorrect label states:
    Digits x 0.1 = mg/l CaCO3 in 100 mL Digits x 0.0002 = meq.
  • The label should state:
    Digits = mg/l CaCO3 in 100 mL Digits x 0.002 = meq
 
Communication to customers
  • The incorrect information on the cartridge label could cause confusion, with the potential for measurement errors if the incorrect information is used. However, the solution concentration and meq calculations are correct, and very few complaints have been received.
  • Please replace cartridges for customers who request this.
  • A letter will sent to affected customers.


Product Information:
 
Affected Lots:
Product # Description Affected Lot #'s
1438901 SULFURIC ACID 1.600N, CARTRIDGE 290, 0351, 0352,1019, 1022, 1046, 1050, 1062, 1076, 1098
 
Kit PNs and Lots Containing Affected DT Cartidges:
Product # Description Affected Lot #'s
2063700 TEST KIT, AL-DT ALK W/DIG TITR 1089B, 1090B, 1094A, 1097A, 1109A, 1089A, 1090A, 1091A, 1095A, 1104A, 0356
2063700-CA TEST KIT, AL-DT ALK W/DIG TITR 1070
2271900 REAGENT SET, ALKALINITY F/DT KIT 1095A, 1098B, 1091A, 1098A, 1104A, 1074, 1038, 1028, 358
2271900-CA REAGENT SET, ALKALINITY F/DT KIT 1064, 1052
2923200 RGT SET, DREL COMPL WATER QUAL 1095A, 1108A
2690200 REAGENT SET, CEL ADV DW 304
243001 rr TEST KIT, FF-2 FISHERIES/FISH 1045
243003 rr TEST KIT, FF-3 FISH FARMING 1067