Can a vortex cooler be used with the C1D2 version of a B3500?

Document ID

Document ID TE12886

Published Date

Published Date 04/14/2021
Print
Question
Can a vortex cooler be used with the C1D2 version of a B3500?
Summary
Vortex Cooler 19-B5C-029 approval for C1D2 B3500 systems
Answer
BioTector's vortex cooler PN: 19-B5C-029 is  designed to be used in hazardous areas.  Part number for BioTector Vortex Cooler
This cooler has ONLY been approved for use with B3500C and B3500ul ATEX systems.  The vortex cooler has NOT been approved for C1D2 versions of the B3500ul or B3500C and therefore should NOT be quoted with a C1D2 system.
An ATEX system uses a TYPE 2 Enclosure ATEX Compliant, European standard, EExp Zone 2 (750mm x 500mm x 320mm).  A C1D2 system uses TYPE 2 Enclosure ETL Certified, North American standard, EExp Class 1 Div 2 (750mm x 500mm x 320mm).  The enclosure type can be determined by the analyzer's part number.