Amperometric titration has USEPA approved methods for chlorine measurement which usually has good correlation with the DPD colorimetric method used with the CL17sc. One needs to ensure the CL17sc is in good operating condition and on the default calibration setting. The grab sample must be representative for both types of analysis and the comparative analysis must be conducted as soon as possible after obtaining the sample.
See the following links and articles for more information on these methods
Chlorine, Free, Amperometric Buret Titration Method 8334_DOC316.53.01155
Chlorine, Total, Amperometric Buret Titration Method 8168_DOC316.53.01156
How do I verify On-Line Chlorine Analyzer using USEPA method 334.0?
You have been directed to this answer based on an ID match.
Click here
to see search results of this phrase.
How does amperometric titration technology compare to CL17sc measurements for chlorine?
Document ID
Document ID TE12807Published Date
Published Date 07/01/2022
Question
How does amperometric titration technology compare to CL17sc measurements for chlorine?
Summary
amperometric titration and CL17sc
Answer