How are results with decimals exported through modbus with an EZ series analyzer?

Document ID

Document ID TE12608

Published Date

Published Date 11/23/2020
Print
Question
How are results with decimals exported through modbus with an EZ series analyzer?
Summary
Explanation of how results with decimals are exported with EZ series analyzers.
Answer
The EZ series analyzers are not able to send decimals through modbus protocols. Instead the result would be multiplied by a factor, usually 100 for common 0-1 ppm applications. The outgoing signal would then need to be divided by the same factor on the receiving equipment in order to restore the result to it's normal form and resolution.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.