Which electrodes can use the feature "live measurement" on the AT1000?

Document ID

Document ID TE12678

Published Date

Published Date 08/13/2020
Print
Question
Which electrodes can use the feature "live measurement" on the AT1000?
Summary
All compatible Intellical electrodes can use the "live measurement" feature, except the double platinum electrode MTC695.
Answer
Select “maintenance” and select “live measure”, you will get the electrode reading in pH/mV.
It works for pH, mV (redox zero current and silver combined electrode, EC (CDC401)) and ISE electrodes.
The option shows continuous measurement data with connected sensors to quickly check measurements. The installed applications and the automatic additions to the sample are set to off. Continuous measurements are not temperature compensated, so measurement differences may occur in the same sample between continuous measurements and measurements that use installed applications with temperature compensation.
 
Live measure feature is not available for the MTC695 double platinum electrode (Imposed current is not managed in that case).
 