Question
What are the min/max values for a custom calibration of the 2100Q turbidimeter?
Answer
The minimum and maximum values which the 2100Q turbidimeter will accept during a custom calibration are:
- 10-40 NTU
- 50-150 NTU
- 400- 1000 NTU