Question
How is the zero calibration enabled on a 5500sc/ 9610sc Silica analyzer
Summary
5500sc/ 9610sc Silica analyzer Zero Cal
Answer
Any software version greater than 1.09 will have the option to perform a zero calibration in addition to the slope calibration.
The zero calibration option can be enabled/disabled in the calibration menu.
This information can be found in the calibration section of the full user manual here.
To initiate a calibration immedately:
The zero calibration option can be enabled/disabled in the calibration menu.
This information can be found in the calibration section of the full user manual here.
To initiate a calibration immedately:
- Press cal to display the calibration menu.
- Select START MANUAL CAL
- Select ZERO CAL if initiating a calibration immediately
- Press cal to display the calibration menu.
- Select SET AUTO CALIBRATION
- Select ZERO CAL