What is the purpose of the sample heater in the 5500/961X Silica/phosphate analyzer?

Document ID

Document ID TE12405

Published Date

Published Date 04/14/2020
Print
Question
What is the purpose of the sample heater in the 5500/961X Silica/phosphate analyzer?
Summary
5500/961X Silica/phosphate sample heater. 
Answer
With the 5500/961X Silica/phosphate analyzers, heating the sample will facilitate the chemical reaction to ensure a fast response time and because the temperature is constant it can ensure accurate results.
The system will heat the sample and the colorimeter to 50° C regardless the incoming sample temperature. If the sample heater is not functioning properly, the analyzer will give a ‘temp out of range’ error. How is the 'Temp Out of Range" Error on a 5500 sc Silica/Phosphate Analyzer resolved?
The heater can be tested using the instructions here:  How test the sample heater on a Series 5500 Silica or Phosphate (120 VAC Models Only).
The heater fuse can be reset with the instructions here: Is it possible to reset the sample heater fuse on the 5500 sc Phosphate and Silica analyzers?

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.