How many 4-20 mA powered outputs are standard for the 5500 silica/phosphate analyzers?

Document ID

Document ID TE12363

Published Date

Published Date 03/04/2020
Question
How many 4-20 mA powered outputs are standard for the 5500 silica/phosphate analyzers?
Summary
5500 4-20 mA Powered outlets
Answer
The 5500/961X silica/phosphate analyzer comes with four 4-20 mA outputs standard regardless of how many sample streams the analyzer is configured with.
The four outputs have their own isolated power.  
If more outputs are needed, additional outputs can be added using the optional output module, 6792200.  Find more information here: Can the sc200 mA output card Product# 9334600 be used with the 5500sc analyzers? .  This is recommended with a system having more than 4 sample streams. There are instructions on configuring this here: Is there an instruction sheet for the output module (Product #6792200) used with the 5500sc Silica and Phosphate analyzers?