What is the air flow path for reagent delivery on a 5500sc or 961Xsc analyzer?

Document ID

Document ID TE12326

Published Date

Published Date 02/12/2020
Print
Question
What is the air flow path for reagent delivery on a 5500sc or 961Xsc analyzer?
Summary
Image and description showing main components for air flow path for reagent delivery on a 5500sc or 961Xsc.
Answer
The labeled image below shows the main components of the reagent air supply and flow indicators. A system which is working correctly will regulate the reagent air pressure from 3.95 to 4.10 psi and typically require the air pump to only come on a couple seconds every one to two hours. It is important for all fittings to be tight and bottle caps snug.





See the following articles for more information on reagent delivery and troubleshooting information: