How can I manually delete a sensor from my SC200 controller?

Document ID

Document ID TE11939

Published Date

Published Date 10/04/2019
Print
Question
How can I manually delete a sensor from my SC200 controller?
Answer
To delete a sensor from your sc200, please follow the below steps:

1. Disconnect sensor from sc200;
2. Power down sc200 (with sensor disconnected);
3. Power up sc200 (with sensor disconnected);
4. You should see a “Scanning for Devices” message displayed during start up, which will clear within 30 seconds;
5. You should then see a “Device Missing” message showing the serial number (or name) of the old sensor;
6. Select ENTER to continue;
7. Select left arrow to select sensor to be deleted;
8. Press ENTER to delete sensor from the sc200.


Sensor is now deleted & a new sensor can be installed.