What are the main specifications of the new XR110 B20B112 reference electrode?

Document ID

Document ID TE11899

Published Date

Published Date 12/20/2021
Question
What are the main specifications of the new XR110 B20B112 reference electrode?
Summary
B20B112 reference electrode XR110, Mercury free reference electrode.
Answer
Main specifications are:
Reference element: Red rod (Ag/AgCl)
Salt bridge solution: KCl 3M
Length: 120 mm
Diameter: 12 mm
Potential versus normal hydrogen electrode E NHE = +209 mV
Temperature range -10, 100°C
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.