When does the relay/LED switch on in case of a low flow at a Filtrax sample unit?

Document ID

Document ID TE11877

Published Date

Published Date 09/16/2019
Print
Question
When does the relay/LED switch on in case of a low flow at a Filtrax sample unit?
Summary
Relay/LED switching to on based on the flow for Filtrax sample unit
Answer
The relay/LED does not switch on not directly when the flow is low for the first time. Both will switch to on if the flow measurement is too low ten times in a row.

Example:  On an instrument, the flow measurement was low seven times, but on the eigth measurement the flow returned to normal.  The flow counter for low flow will now start again at zero.