Is there a way to connect the sc200 to the sc1000 in order to read the values of the sc200 connected sensors on the sc1000 display?

Document ID

Document ID TE11867

Published Date

Published Date 01/05/2022
Question
Is there a way to connect the sc200 to the sc1000 in order to read the values of the sc200 connected sensors on the sc1000 display?
Summary
Customer intension is to share information of probes that are connected to sc200 into sc1000 controller.
Answer
Yes to read information from probe that are connected to sc200 on sc1000 controller. 
Customer has to use mA output card in sc200 controller and scale measurement values of probe with mA output card. An then install mA Input card in sc1000 and connect this mA input card to mA output from sc200. This will permit you to get measurements values of sc200 onto sc1000 display.