What is the acceptance range for the calibration slope of LDO101 probe?

Document ID

Document ID TE11258

Version

Version 2.0

Status

Status Published

Published Date

Published Date 06/20/2019
Question
What is the acceptance range for the calibration slope of LDO101 probe?
Summary
Acceptance range for the calibration slope for the LDO101.
Answer

There is no acceptance range for the LDO101 probe; there is no calibration slope as such but rather a calibration reference point at 100% in humidity saturated air (~21% O2 content by volume of air), so once dipped in water the meter calculates the theoretical value of dissolved oxygen at specific temperature, pressure & salinity.

 

 