What is the recommended storage for an IntelliCAL MTC695 Double Platinum Electrode in an application for Chlorine, Chlorine Dioxide, and/or Chlorite?

Document ID

Document ID TE11227

Version

Version 1.0

Status

Status Published

Published Date

Published Date 05/30/2019
Question
What is the recommended storage for an IntelliCAL MTC695 Double Platinum Electrode in an application for Chlorine, Chlorine Dioxide, and/or Chlorite?
Summary
Storage instructions for an MTC695 used for Chlorine testing.
Answer
For short term storage (Overnight/Over the weekend), store the electrode in 1% Nitric Acid in chlorinated tap water (alternatively DI water with a drop of bleach would be fine too).
For long term storage (a week or longer), the electrode can be stored dry. Always clean using the programmed procedure before use. Multiple cleanings may be required after extended dry storage.

Here is an article with the instructions on this cleaning that would be needed after dry storage: What are the cleaning instructions for the IntelliCAL MTC695 Double Platinum Electrode for AT/KF1000 RedOx Titrations?