What can be verified on a FL90x with a detected modem that is missing calls?

Document ID

Document ID TE11179

Version

Version 1.0

Status

Status Published

Published Date

Published Date 05/23/2019
Question
What can be verified on a FL90x with a detected modem that is missing calls?
Summary
FL90x communication troubleshooting
Answer
If a FL90x is missing calls, first verify that the modem is being detected within the FSDATA Desktop under the communication tab:

If the modem is recognized then the following can be verified:
- The Antenna installed properly
- There adequate strength (-113 worst, -50 best)
- That the IP address is correct
- That the SVC Code is correct
- Verify that FSData Online is setup with the loggers serial number and added to the correct SVC location.
- Verify that the SIM card number is recognized after the FL900 is power cycled under modem diagnostics. 
- Verify that the SIM card is inserted correctly: 
An incorrectly installed SIM is seen below:

A correctly installed SIM is seen here: