Is cleaning of the NV3300 sensor optics available?

Document ID

Document ID TE11054

Version

Version 2.0

Status

Status Published

Published Date

Published Date 03/21/2019
Question
Is cleaning of the NV3300 sensor optics available?
Summary
Cleaning systems for the NV3300 sensor
Answer
Yes. The cleaning system uses compressed air and can be provided at the site (with the customer’s compressed air source fed directly into the CD500 controller) or by our new HPAC (High Pressure Air Cleaning) cleaning system (a compressed air unit, like Hach’s HOAB). The wiper clean is not suitable in combination with the NV3300 due to the long optical path length of these sensors.