How are the spiking rates for LCK kits using Addista LCA Standards determined?

Document ID

Document ID TE10962

Published Date

Published Date 04/16/2019
Question
How are the spiking rates for LCK kits using Addista LCA Standards determined?
Summary
Spiking rate for Addista Standards
Answer
The standard solution of the addista kits can be used as spiking solutions as well. On the back side of the Addista box customers can find all parameters/LCK test kits which can be analyzed with the respective kit. The customer finds the nominal concentration of the standard and next to this the spiking rate. The spiking rate is always half the value of the standard, for example, if the standard has a concentration of 3 mg/l, the spiking rate is 1.5.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.