What should be the distance between Solitax sc and wall?

Document ID

Document ID TE10960

Published Date

Published Date 04/17/2019
Print
Question
What should be the distance between Solitax sc and wall?
Summary
Distance between Solitax sc and wall
Answer
As stated in the SOLITAX® sc Sensor User Manual the distance between sensor and wall should be:
- TS > 10 cm (4 in)
- Turbidity >50 cm (20 in).

Beyond that a guideline can be used to be more specific:
Range:                         Distance to the wall:
    >10FNU                     >  10 cm (colour of the wall don't matter)
    >  1FNU                     >  10 cm (colour of the wall: black)
    >  1FNU                     >  30 cm (colour of the wall don't matter)
    >  0.1FNU                  >  30 cm (colour of the wall: black)
    without limitation     >  60 cm (colour of the wall: black)
    without limitation     >100 cm (colour of the wall don't matter)
 