Why the Data Log of DR3900 spectrophotometer shows more as 2000 results?

Document ID

Document ID TE10934

Version

Version 2.0

Status

Status Published

Published Date

Published Date 04/23/2019
Question
Why the Data Log of DR3900 spectrophotometer shows more as 2000 results?
Summary
Data Log DR3900 spectrophotometer
Answer
The technical specification in the user manual describes the Data Log with a capacity of 2000 readings. The instruments show 2498 readings (see screenshot).


 

The Data Log has a minimum reading of 2000. The AQA Log has a minimum reading of 500.
If the AQA Log is empty the Data Log will use this empty space in addition to the 2000 readings (2000 + 500 = 2500). Therefore it can be that the instrument shows up to 2500 readings in the Data Log.