Do Hach have a suggested conductivity range for the ISE sc sensors?

Document ID

Document ID TE10925

Version

Version 2.0

Status

Status Published

Published Date

Published Date 04/24/2019
Question
Do Hach have a suggested conductivity range for the ISE sc sensors?
Summary
Conductivity range for ISE sc sensors
Answer
The typical environment for ANISE is biological phase of the municipal treatment plant. Unfortunately, Hach does not have a suggested conductivity range.
It is still meaningful using an ISE-sc sensor in drinking water with 600/700 μS/cm. The ISE sc sensors were not tested in milieus with lower conductivity concentration.
Using ISE sc in water with lower conductivity could be critical for the salt bridge and lead to measurement drifts.
Customers may test the sensor on the site, but Hach cannot guarantee its functionality.