Does the weld-on fitting with tri-clamp connection for TSS probes come with a pre-shaped end?

Document ID

Document ID TE10814

Published Date

Published Date 09/16/2019
Print
Question
Does the weld-on fitting with tri-clamp connection for TSS probes come with a pre-shaped end?
Answer
Yes, the weld-on fitting (Product # LZU302.99.00000) is available in different variants depending on the size of the pipe which it has to be mounted on.

It is suitable for all TSS tri-clamp sensors except EX1, Titanum, VARI and XL models.

This fitting is not recommended when working with the smallest connection DN65, measuring low turbidity because interference from reflections of the opposite side of the pipe can disturb the measurement.

Installation instructions can be downloaded here.