What is the flow rate of the AT1000/KF1000 pumps to determine volume of reagent addition in a method?

Document ID

Document ID TE10724

Version

Version 1.0

Status

Status Published

Published Date

Published Date 11/01/2018
Question
What is the flow rate of the AT1000/KF1000 pumps to determine volume of reagent addition in a method?
Summary
Specification of flow rate for the titrator pumps and an explanation on how to use that information to determine the volume of reagent added by pump.
Answer
The flow rate for the AT1000 pumps are approximately 100 mL/min.
The flow rate for the KF1000 pumps are approximately 170 mL/min.

These specifications can be used to determine what volume of reagent or solvent is added when the pumps run.

For example if the pump runs for 3 seconds on an AT1000: (100/60)*3 = 5 mL of reagent.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.