Question
What is needed to remotely start an AS950 sampler?
Summary
Remote start of an AS950 sampler
Answer
To remotely start an AS950 Sampler Controller of the following wiring options are needed:
- 8528500 AUX Cable - HALF, 4 ft.
- 8528501 AUX Cable - HALF, 25 ft.
- 9501000 Auxilary junction box

Wiring to be performed in accordance to table below - Pin 2 & 4

cid:image002.png@01D45639.50D5CAE0

Select Hardware Setup>AUX and I/O Port.

Make sure that Type is set to AUX
Go to Programming>Sample Programming>Program Start>On Trigger - Select External AUX
Enter Sampling routine