What are the expected resistance values of the wires of a 3700 Series Inductive Conductivity Sensor?

Document ID

Document ID TE10645

Version

Version 1.0

Status

Status Published

Published Date

Published Date 10/25/2018
Question
What are the expected resistance values of the wires of a 3700 Series Inductive Conductivity Sensor?
Summary
Resistance values for the wires of a 3700 Series Inductive Conductivity Sensor.
Answer
When troubleshooting a 3700 Conductivity Sensor the following pairs of wires should produce the associated resistances:
Between red and yellow wires:  1090-1105 ohms at 23-27 C
Between blue and white wires:  Less than 5 ohms
Between green and yellow wires:  Less than 5 ohms
Between white and shield wires:  Infinite (open circuit)