What are the low flow and high flow options on the CL17 sample conditioning kit?

Document ID

Document ID TE10525

Version

Version 1.0

Status

Status Published

Published Date

Published Date 10/25/2018
Question
What are the low flow and high flow options on the CL17 sample conditioning kit?
Summary
Explanation of the low flow and high flow connections on the CL17 sample conditioning kit mentioned on page 17 of the user manual.
Answer
The low flow and high flow options listed on page 17 of the user manual are two different connections for the start of the sample conditioning kit.  The low flow option has 1/4 inch tubing and the high flow option has 1/2 inch tubing.  The decision of which connection to use depends on customer preference and application, however the psi for both connections has to be at or below 87 psi.