Question
Are there known limitations on AS950 I/O box (IO9001 & IO9004)?
Summary
Known limitations on AS950 I/O box (IO9001 & IO9004)?
Answer
IO9001
Main feature of the IO9001 is 1x AC output relay. The digital inputs are disabled – digital flow input won't work - if relay output and flow input should be used at the same time, the flow signal must be analogue
IO9004
The IO9004 has 4x AC output relays, 2x analog inputs, 3x analog outputs, and 4x digital outputs. The digital flow inputs are disabled – digital flow input won't work - if relay output and flow input should be used at the same time, the flow signal must be analogue
Main feature of the IO9001 is 1x AC output relay. The digital inputs are disabled – digital flow input won't work - if relay output and flow input should be used at the same time, the flow signal must be analogue
IO9004
The IO9004 has 4x AC output relays, 2x analog inputs, 3x analog outputs, and 4x digital outputs. The digital flow inputs are disabled – digital flow input won't work - if relay output and flow input should be used at the same time, the flow signal must be analogue