How long will the FL1500 Battery Backup last?

Document ID

Document ID TE10097

Published Date

Published Date 07/03/2018
Question
How long will the FL1500 Battery Backup last?
Summary
FL1500 DC Battery Backup
Answer

The FL1500 draws 15 watts of power. 
If we want the unit to run for X hours, the math would be  15 watts * X hours = Total watt hours. Batteries are rated in amp-hours, so to calculate the amp-hours needed you would divide the Total watt-hours, by 12 volts to get the needed amp-hours.
The DC battery backup Hach ships is 7 amp-hours, this battery would power the logger for approximately 6 hrs.

Please note that these are estimates and actual battery life may vary.