Why is there no option within the controller to perform a dry verification on the Filtertrack 660 (FT660)?

Document ID

Document ID TE9953

Published Date

Published Date 06/07/2018
Print
Question
Why is there no option within the controller to perform a dry verification on the Filtertrack 660 (FT660)?
Summary
FT 660 dry verification
Answer
If the calibration verification module (CVM) is not enabled within the controller, you will not have an option to perform a dry verification within the sensor menu. Please follow the instructions below to enable this option.

1. From the Main menu, select Sensor Setup, [Select Sensor], Calibrate, Verification, CVM Select, Enable.
2. To set the pass or fail criteria to a value other than the recommended default ( ±  5 mNTU):
     a. Select Verification, P/F Criteria.
       Note: The P/F Criteria menu option is available only if CVM is set to ENABLE in step 1.
     b. Use the arrow keys to change the PFC (pass or fail criteria) value a nd confirm.
       The PFC range is 5 - 250 mNTU