What are the calibration limits for the 5500 sc Silica and Phosphate analyzers?

Document ID

Document ID TE9789

Version

Version 1.0

Status

Status Published

Published Date

Published Date 05/07/2018
Question
What are the calibration limits for the 5500 sc Silica and Phosphate analyzers?
Summary
Calibration limits for the 5500 sc Silica and Phosphate analyzers
Answer
Below are the calibration limits for each of the instrument configurations.


Additional information:
5500 sc Silica
Measurement Range: 0.5-5000 μg/L as SiO 2
Accuracy: 0-500
μg/L: ±1% or ±1 μg/L (whichever is greater), 500-5000 μg/L: ±5%
Precision/Repeatability: ±0.5 ppb or ±1% (whichever is greater)

5500 sc Low Range Phosphate
Measurement Range: 4-3000 μg/L as  PO 4
Accuracy: ±4 μg/L or ±4% (whichever is greater)
Precision/Repeatability: ±1%

5500 sc High Range Phosphate

Measurement Range: 200-50,000 μg/L as  PO 4
Accuracy: ±500 μg/L or ±5% (whichever is greater)
Precision/Repeatability: ±500 μg/L or ±5% (whichever is greater)

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.