How is a network created between SC1000 controllers?

Document ID

Document ID TE9719

Published Date

Published Date 06/05/2019
Print
Question
How is a network created between SC1000 controllers?
Summary
SC1000 network
Answer
Note: This article addresses creating an internal BUS network. For information about creating external network connections to a SCADA or PLC, please see the sc1000 Controller Enhanced Communications Manual.

When creating an SC1000 controller network, you will need one SC1000 Internal Network Connector (Product # LZX918) per SC1000 probe module as well as a run of BUS cable (Product # LZY488 - indoor and flexible or Product # LZY489 - outdoor and rigid) which is sold by the foot. A network can be composed of up to 32 devices. . There can only be one display module per SC1000 network and it does not matter which probe module the display module is installed on.



Connecting the network cable to the SC1000 probe modules:
The outside connectors (represented above by Probe Module 1 and Probe Module 3) should always be connected to 1A and 1B and the terminating switch should be turned on.  This places a resistor (~220 Ohms) across the 1A and 1B connection and also disables the 2A and 2B connection making it inoperable.  The center probe module should have the two cables connected to it, one on 1A and 1B and the second on 2A and 2B.  The switch should be in the off position to disable the terminating resistor and enabling connection 2.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.