What are the default calibration standard values on the APA 6000 High Range (HR) Hardness Analyzer?

Document ID

Document ID TE9703

Version

Version 1.0

Status

Status Published

Published Date

Published Date 04/10/2018
Question
What are the default calibration standard values on the APA 6000 High Range (HR) Hardness Analyzer?
Summary
APA 6000 High Range Hardness calibration standard values.
Answer

On the APA 6000 HR Hardness analyzer, standards 1 and 2 are used together to perform a multiple-point calibration. Standard 1 is a 0 mg/L as CaCO 3 sodium carbonate solution and standard 2 is a 1000 mg/L as CaCO 3 sodium carbonate solution. The default calibration standard values that are created by the analyzer are 0 mg/L, 100 mg/L, 500 mg/L, and 1000 mg/L. 
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.