Can the SP510 Hardness Analyzer send a 4-20 mA output signal?

Document ID

Document ID TE9661

Version

Version 1.0

Status

Status Published

Published Date

Published Date 05/08/2018
Question
Can the SP510 Hardness Analyzer send a 4-20 mA output signal?
Summary
SP510 Outputs
Answer
The SP510 Hardness Analyzer does not have a 4-20 mA output option. This instrument does not actually measure hardness, but triggers a relay based on a trip point determined by different reagent/buffer combinations. If a customer needs to measure exact hardness values and have a 4-20 mA output proportional to the measured value, then they will need an APA 6000 Low Range Hardness Analyzer (Product # 5100210) or an APA 6000 High Range Hardness analyzer (Product # 6200010).