Question
Do IntelliCAL™ probes need to be recalibrated if disconnected from a meter?
Summary
Calibration memory on IntelliCAL™ probes.
Answer
IntelliCAL™ probes store calibration data, calibration history, and probe settings on the probes themselves. This is in contrast to most other meters and probes which store the calibration data and history on the meter instead of the probes.

Because of this an IntelliCAL™ probe can be calibrated on one meter, disconnected and connected to a second meter, and continue to use the calibration that was performed while the probe was still connected to the first meter.

In summary IntelliCAL™ probes do not require recalibration when disconnected from the meter.