How to connect a liquid level switch to a BÜHLER (BUHLER, BUEHLER) 1027 or 2000 sampler?

Document ID

Document ID TE9528

Version

Version 1.0

Status

Status Published

Published Date

Published Date 03/13/2018
Question
How to connect a liquid level switch to a BÜHLER (BUHLER, BUEHLER) 1027 or 2000 sampler?
Summary
Connection of a liquid level switch to a BÜHLER (BUHLER, BUEHLER) 1027 or 2000 sampler
Answer
To connect a liquid level switch ( Liquid Level Actuator ) to a BÜHLER 1027 or 2000 sampler the follwing parts are needed:
- Liquid Level Actuator - order code / article number 943
- External Signal Connector -
order code / article number BM10064
- Alternative to BM10064 also the 10m signal cable with connector can be used order code / article number BM69644

Connector of Liquid Level Actuator needs to be cut off. The 2 wire of Liquid Level Actuator needs to be connected to PIN 3 and 5 of the External Signal Connector plug.

Finally in the sampler programming the sampling mode needs to be changed to Event / time based sampling.
 