How can you perform a blank measurement with LCW058?

Document ID

Document ID TE9520

Version

Version 1.0

Status

Status Published

Published Date

Published Date 03/13/2018
Question
How can you perform a blank measurement with LCW058?
Summary
Zero/ Blank measurement in LCW058 Hydrogen peroxide pipette test 1-10 g/L H 2 O 2
Answer
The user has to insert the full and closed Standard cuvette to zero the instrument. The Standard cuvette works in this test kit the as the zero cuvette.
The content of this cuvette will not to be used as a standard solution in the test.
It just contains the needed zero solution and has not to be opened.
In this test kit we do not deliver a standard solution for the customer.
 
So to make it clear.
The user has to insert the Standard cuvette, press zero and then measure the prepared sample cuvettes (LCW 906) and press read.
After this they will get the result.
If the user wants to measure a standard they have to prepare a standard solution by themselves.
 