How to calibrate the U53 Sensor on a sc200?

Document ID

Document ID TE9385

Published Date

Published Date 10/25/2018
Question
How to calibrate the U53 Sensor on a sc200?
Summary
Procedure for calibrating U53 Sensor using two-point calibration.
Answer
  1. Menu > Sensor Setup > Calibrate
  2. Select 2 pt Depth Cal, enter
  3. Select Active, enter
  4. Wait for sensor to range stabilize, enter
  5. Measure the distance from sensor face to surface water and enter value (set Range), enter
  6. Wait for sensor depth to stabilize, enter
  7. Measure the depth of the water under the sensor and enter value (set Depth), enter
  8. Will Display Passed or Failed & Select if a new sensor or not
  9. Calibration complete