What is the maximum cable distance from SC1000 Display Module to Probe Module?

Document ID

Document ID TE9317

Version

Version 1.0

Status

Status Published

Published Date

Published Date 03/01/2018
Question
What is the maximum cable distance from SC1000 Display Module to Probe Module?
Summary
Maximum distance between SC1000 Display Module and Probe Module
Answer
The SC1000 display cable can be extended up to 15 meters.  The display will not power up due to power loss when longer cables are used. 
Hach offers the following extension cables that can be used between the SC1000 Probe Module and Display Module:
LZX847: 0.35 m sc-extension cable for sc100/sc1000
LZX848: 5 m sc-extension cable for sc100/sc1000
LZX849: 10 m sc-extension cable for sc100/sc1000
LZX850: 15 m sc-extension cable for sc100/sc1000