What is the accuracy and resolution of Radiometer burettes?

Document ID

Document ID TE9315

Version

Version 1.0

Status

Status Published

Published Date

Published Date 02/16/2018
Question
What is the accuracy and resolution of Radiometer burettes?
Summary
Resolution and accuracy for Radiometer burettes.
Answer
The Radiometer titrators are capable of 18,000 steps. Depending on the burette volume this can be used to calculate the resolution. For example a 10 mL burette like the B510 (Product # X51T017) is equivalent to 10,000 µL. 10,000 µL/ 18,000 steps = 0.55 µL/step.

These burettes are manufactured according to ISO 8655-3 specifications. Accuracy information for the burettes can be found at the end of the instrument reference manual for the titraor (Appendix 4 -- Technical Specifications). Below is from this section of the manual.