How do you take a diagnostic reading using the Flo-Dar with the FL900 in FSData Desktop?

Document ID

Document ID TE9168

Version

Version 2.0

Status

Status Published

Published Date

Published Date 10/25/2018
Question
How do you take a diagnostic reading using the Flo-Dar with the FL900 in FSData Desktop?
Summary
How to take a diagnostic reading using the Flo-Dar with the FL900 in FSData Desktop.
Answer
  1. Open FSData Desktop and connect to the FL900 Logger.
  2. When you connect to the FL900, there is a tall, narrow section at the far right called "FL900 Status."  There are two buttons there at the top.
    • Current Status
    • Diagnostics
  3. Click on the "Diagnostics" button.
  4. At the top right there is a box labeled "Sensor."
    • If you only have one sensor, it will be in the box
    • If you have two sensors, make sure that the sensor you want to take a sample with is in the box.  If it is not, click on the triangle on the right and select the sensor you want.
    • If your sensor is not listed:
      • Check the cable connections.
      • If that doesn't help, reboot the FL900 Logger.
      • If that doesn't help, reflash the firmware on the FL900 Logger.
      • If that doesn't help either the FL900 or the FLO-DAR is not working.
  5. Below "Sensor" is the "Take Measurement" button.  Click on it.
    • When you click on it, the button will grey out.
    • At the far bottom, left you will see a message that says "Waiting for sensor data…XX", where XX is the number of seconds since the reading started.
    • A reading typically takes about 50 seconds, but can sometimes take longer.
  6. When the reading is finished the data will be displayed below the "Take Measurement" button. 
    • At the bottom there will be the velocity graph.
    • To see an enlarged version of the graph with the sensor data displayed, click on the square button to the right of the "Take Measurement" button 