What is the NOL programming parameter for the Flo-Dar with the FL900?

Document ID

Document ID TE9164

Version

Version 2.0

Status

Status Published

Published Date

Published Date 10/05/2018
Question
What is the NOL programming parameter for the Flo-Dar with the FL900?
Summary
Understanding the NOL programming parameter for the Flo-Dar with the FL900
Answer
  1. NOL
    • NOL stands for Number of Levels.
    • NOL is related to the velocity measurement only. It’s purpose is to weed out bad velocity readings.
    • If a bad velocity reading is obtained (meaning Vel is set to 0), an algorithm will take the last good [Level, Velocity] reading and create a new velocity reading. 
    • If we have (L1, V1) and (L2, V2) where V2 was defaulted to 0, then the new reading V2 = (L2/L1)*V1.
    • NOL is the number of times this will be done.  If NOL=5, then it will do this 5 times.  After that, the velocity will be set to 0.
    • Default=5