What are the different Flo-Dar surcharge programming parameters on the FL900?

Document ID

Document ID TE9163

Version

Version 2.0

Status

Status Published

Published Date

Published Date 10/26/2018
Question
What are the different Flo-Dar surcharge programming parameters on the FL900?
Summary
Understanding the Flo-Dar surcharge programming parameters on the FL900
Answer
  1. Conductivity Trigger.
    • This is the value reading off of the SVS conductivity sensor that signals that the SVS is touching water.  The units are micro-Siemens/cm.
    • Default=50
  2. Distance Turn on and Distance Switchover
    • When the ultrasonic sensor's distance reading reads less than 8" (meaning there is water 8" from the ultrasonic sensor), the SVS sensor will turn on.
    • When the ultrasonic sensor's distance reading reads less than 7" (meaning there is water 7" from the ultrasonic sensor), the Flo-Dar will "switchover" from the main sensors to the SVS sensor.  The radar velocity and ultrasonic level sensors will turn off.
  3. Surcharge Level Trigger
    • The Flo-Dar has a pressure level sensor that is used for surcharge detection and to measure the depth of flow during surcharge.
    • When the Surcharge Level Trigger is exceeded, the level will read full pipe.
    • Default = 1".
  4. SVS Velocity Multiplier
    • This a calibration adjustment to make the SVS velocity match the actual velocity.  The velocity measured by the SVS is multiplied by the number in the box.
    • For example, if the SVS measures a velocity of 2.0 fps and the actual velocity of the water is 2.4 fps, then enter 1.20 into the Site Multiplier box.
  5. What makes the Flo-Dar switchover to the SVS sensor?
    • There are three parameters monitored:
      • Surcharge Level
      • Conductivity
      • Distance
    • Whenever two of these three parameters indicate surcharge, the Flo-Dar will switchover to the SVS sensor.  If only one of them indicates surcharge, nothing will happen.
  6. These parameters are somewhat obscure.  Do not change them unless you understand them well.  If you have questions about these call Hach Flow Tech Support at 800 368 2723 ext. 6063.