Is there any support data available for the Ozone Method 8311 in place of a missing Method Performance Section?

Document ID

Document ID TE9137

Version

Version 2.0

Status

Status Published

Published Date

Published Date 07/31/2018
Question
Is there any support data available for the Ozone Method 8311 in place of a missing Method Performance Section?
Summary
Support Data for Ozone Chemistry.
Answer
The Ozone Method 8311 does not have a Method Performance section like other methods. Below and attached is support data for this chemistry.

Method Detection limit*
Ozone Test Average (mg/L) MDL (mg/L)
Low Range (0.01-0.25 mg/L) 0.02 0.022
Mid Range (0.01-0.75 mg/L) 0.05 0.034
High Range (0.01-1.50 mg/L) 0.03 0.026
*Seven replicates were analyzed

Precision Report*
Ozone Test Average (mg/L) %RSD
Low Range (0.01-0.25 mg/L) 0.128 13.4
Mid Range (0.01-0.75 mg/L) 0.39 2.09
High Range (0.01-1.50 mg/L) 0.54 3.82
*Four replicates were analyzed
 