Why will my CL10 chlorine sensor not calibrate after changing the electrolyte?

Document ID

Document ID TE9061

Published Date

Published Date 01/06/2022
Print
Question
Why will my CL10 chlorine sensor not calibrate after changing the electrolyte?
Summary
CL10 chlorine sensor cap loose, --.-- reading low
Answer
The most common problem with the CL10 not calibrating or reading low is that the membrane cap is not on tight enough. Please turn the cap until it actually stops turning. This brings the membrane closer to the electrode and increase the analytical response.