What is the dilution water referenced in the TU5200 Laboratory Laser Turbidimeter user manual?

Document ID

Document ID TE9052

Version

Version 3.0

Status

Status Published

Published Date

Published Date 10/18/2018
Question
What is the dilution water referenced in the TU5200 Laboratory Laser Turbidimeter user manual?
Summary
TU5200 Laboratory Laser Turbidimeter dilution water
Answer
The dilution water referenced in the TU5200 Laboratory Laser Turbidimeter user manual should be prepared by filtering deionized water through a 0.20 micron filter and degassing (using the Sample Filtration and Degassing Kit, Product # 4397510 or equivalent) to ensure that the turbidity is less than 0.5 NTU.
 