What are specifications for the analog outputs for the FL1500?

Document ID

Document ID TE8682

Published Date

Published Date 02/19/2019
Print
Question
What are specifications for the analog outputs for the FL1500?
Summary
Analog output specifications for the FL1500
Answer

Output options for the FL1500 include two (basic model) or three (advanced model) 0/4–20 mA outputs (0/4–20 mA current loop; external loop power 18 VDC maximum or internal loop power 14 VDC minimum, 18 VDC maximum; loop burden 3.6 V maximum at 25 mA).

Maximum cable length depends on several factors including:
* The loop power supply voltage. The customer may choose to use the FL1500 internal (specified as 14V minimum) or an external (specified as 18V maximum) power supply.
* Wire gage of the “loop” cable.
* Loop load resistance. This is the resistance of the load device that the customer connects to the current loop. The customer must supply this information.
* FL1500 loop burden, specified as “3.6V max @ 25mA”
* The maximum loop current. For the FL1500 the maximum is 25mA.

Please see the attached document for more information.