What is the proper way to check the memory battery voltage and memory battery status on the Flo-Station?

Document ID

Document ID TE8664

Version

Version 3.0

Status

Status Published

Published Date

Published Date 08/21/2018
Question
What is the proper way to check the memory battery voltage and memory battery status on the Flo-Station?
Summary
Checking the memory battery voltage and memory battery status on the Flo-Station.
Answer
Checking the battery voltage and battery status:
Click 'Real Time' and then Click 'Sample'.  Once real time data populates (~45seconds) click the down arrow on any of three channel dialog boxes and select 'Backup Battery'.
     -3.30-3.70vdc= battery has sufficient voltage
     -vdc<3.20= battery needs to be replaced.

Also see Backup Battery Status:  Poor = replace battery  Good = battery has sufficient voltage



For more information on the Memory Battery, please refer to What is the internal memory backup battery inside the Flo-Station?
 