Question
How do you perform a Zero Adjustment on the Flo-Mate 2000?
Summary
Performing a Zero Adjustment on the Flo-Mate 2000
Answer

Before completing the following steps, perform a Zero Velocity Check first.

Zero Adjust

  • Position the sensor as described in the zero check procedure. 
  • To initiate the zero start sequence, press the STO and RCL keys at the same time. You will see the number 3 on the display.
  • Decrement to zero with the Down Arrow key.


  • The number 32 will be displayed.
  • The unit will decrement itself to zero and turn off. The unit is now zeroed.

NOTE: Each key in the zero adjust sequence must be pressed within 5 seconds of the previous key. If the time between key entries is longer than 5 seconds or if a wrong key is pressed, the unit will display an ERR 3. Turn the unit OFF then back ON and try again.